Model TS01-16
Miniaturized Thermocouple Scanner

|
Product Description
The all new miniaturized multichannel thermocouple scanner comes with a handful of features in a compact ruggedized enclosure. Developed as part of an integrated data acquisition system, the thermocouple scanner comes with 16 input channels and built+-in cold junction compensation. Screw type terminals accessible from the top make it easy to connect the thermocouples. All 16 channels share a single signal conditioning path using time shared multiplexing controlled by the master controller. A high speed 18 bit ADC digitizes the data and transfers the data stream to an Ethernet port in real time. |
SPECIFICATIONS
Thermocouple:
- Supports B, C, E, J, K, N, R, S, and T type thermocouples
- Internal cold junction compensation
Input specs:
- ±100 mV range
- Differential, 2 wire grounded or ungrounded thermocouples
- 5 KΩ input impedance matching with typical thermocouple impedance range
Calibration:
- Internal reference voltage input to differential amplifier.
- Amplifier inputs short mode for offset calibration.
Amplifier specs:
- Gain of 20 (Fixed)
- Gain stable to 0.02%, ±2 ppm/ 0C
ADC specs:
- 18 bit resolution, serial data bus
- 570 kSPS data throughput
- Linearity ±2.5LSB, Gain error ±.048% FSR
System specs:
- Compact design in Metal Case (120 x 80 x 60 mm)
- Simplified Ethernet interface implemented with single board computer
- Scan rate of 200 samples/ channel/ second
- Uniform Reference Junction temperature compensation to within +/- 0.2 oC using RTD
- Built-in Thermocouple break detection, auto zero and channel offset compensation
- PC software with Thermocouple test, single capture, and continuous read modes. Data displayed as microvolt readings.
- 0.8 μv resolution on -100 to +100 mv input (18-bit)
- Channel Noise level: 10 μv rms
- Operating temperature: -10 to +65 oC
- Power supply Nominal: 12 to 28 DC volts (12V 220 mA Typ.)
- Internally fan to minimize temperature gradient buildup.
|
|